.. _version6:
.. |br| raw:: html
.. include::
ADU-07e Version6 HTTPS
========================
The Version 6 release is compatible to the ADU-08e. |br|
The XML jobfiles as well as structure of the database are the same as for ADU-08e. |br|
Also the version 6 enables the **tablet & mobile phone** operation by the :doc:`../app/app` which was introduced with the ADU-08e. |br|
Version 6 is the *last* version for ADU-07e, and only updated in case of bugs detected.
All previous versions are discontinued. |br|
You need a **USB WiFi dongle** to use WiFi and / or the tablet. The internal WiFi is off (transmits too much into the box). |br|
You have to login via **https** remind that!!
A dongle which works is the :ref:`LogiLink ` .
Download the `ADU-07e system software `_
Documentation for older systems can be found here `ADU-07e and older systems `_
Benefits
----------
Mobile phone and tablet operation. |br|
Joblists for :doc:`../insitucoilcalibration/insitucoilcalibration` and parallel sensor test. |br|
Hassle free operation with ADU-08e together. |br|
Newer software such as *CoilFieldTestTool* (for evaluating parallel tests) and *InSituCoilCalibrationTool* (generate calibration functions with the build in calibration windings
and compare with the manufacturers calibration) will only run with *V6* software.
Remarks for ADU-07e Systems
--------------------------------
.. index::
single: Radio Filter 07e
ADU-07e Radio Filter
----------------------
.. csv-table:: adu07e_radio_filter
:header: "name", "alias", "JLE alias", "contact resistance", "value", ADU-08e filter, "comment"
:delim: ,
RF_1 , LF_RF_1 , 3000 , > 3000 |OHgr| , 22 pF, ,same as below
RF_2 , LF_RF_2 , 1000 , < 3000 |OHgr| , 122 pF, 470 pF; 3000 **RF_1** ,resistive environment
RF_3 , LF_RF_3 , 500 , < 1000 |OHgr| , 6822 pF, 6800 pF; 1000 **RF_2** ,conductive environment
RF_4 , LF_RF_4 , 200 , < 500 |OHgr| , 6922 pF, ,same as above and ***BUFFER** cables
**RF_3** is the default for conductive ground. |br|
RF_4 you only take at very conductive and buffer electrodes (but also RF_3 works here).
RF_1 similar to RF_2 |br|
**RF_2** is for dry and resistive.
If you set the LP_4Hz low pass filter, take RF_1. The low pass makes sense only when using a fluxgate. |br|
(On ADU-08, if radio filter is in off position (e.g. HF measurement) it will automatically switched to RF_2 when measurement is set to LF).
.. image:: /media/adu-07e/adu07e_radio_filter.png
:width: 80%
:align: center
Selftest Warning for ADU-07
----------------------------
For ADU-07 (without e !) you get at selftest warning : |br|
**ADU-07-CON Warning unknown connector type (maybe old Rev. 1.0)** |br|
That is interpreted as possible failure because the ADU can not see the possibilty to connect to the coil. |br|
Your ADB error should be "0". |br|
On the screen shot you see the laboratory situation - no E connected and ADB error 36 - *probe resistance out of range - check electrode and cable*
... no wonder
.. image:: /media/adu-07e/warning_old_con_board.png
:width: 80%
:align: center