Insitu Coil Calibration
===================================
The Insitu Coil Calibration can be performed in the field. |br|
It is an additional tool for checking the calibration data of the induction coils of the MFS serires. |br|
The major difference between a parallel test and a coil calibration check is that the calibration signal is much stronger compared to the natural field. |br|
That results to that it can happen that the calibration check still looks good, however the :ref:`parallel test ` shows a difference. So *both* tests must be performed in order to
check the complete functionality of the MFS coil.
E-Field must NOT be connected! (but will be recorded) |br|
Since we feed in a strong signal, the signal amplitude may *exceed* the maximum level at *noisy* sites. |br|
Or simply spoken: at locations where you do MT it will work; at locations inside a city (or your laboratory) it will not work. |br|
You can download the corresponding joblist
and data here: `Insitu Coil Calibration `_.